As a professional UV light meter manufacturer, we have developed several UV radiometers for a variety of applicable industries during the last 16 years since the foundation of Linshang. Boasting a leading research and development team and a mature, stable quality management system, Linshang has the confidence that every UV light meter can meet the user's needs.

According to the ultraviolet wavelength (UV-A, UV-B, UV-C, UV-V) and the type of UV lamp (high pressure mercury lamp, LED light source, etc. ), our UV light meters can be divided into a variety of models. 

86125 Light Meter+BL-X0 Probe

1.Suitable for blue light energy and power detection

2.Spectral response range: 410nm-490nm

3.Power measuring range: 0 - 2000mW/cm²

4.Advanced digital probe technology, it has high precision and good anti - jamming ability

Browse:27

86129 Ultraviolet Digital Probe

1.Communication protocol: MODBUS;Communication interface: RS485
2.Spectral response: 320nm-400nm;λp =365nm
3.It is convenient to communicate with PLC, human-computer interface, computer and other equipment and collect and monitor real-time data.

Browse:58
prev 12 Next 2 pages 17 items

How to choose a suitable UV light meter?

Different wavelength ultraviolet light sources have different functions. For example, UVC wavelength ultraviolet light is generally used in the sterilization industry and UVA wavelength ultraviolet light is used in the curing industry. When choosing an UV light meter, please firstly confirm the ultraviolet light source you need to measure -type or band. This information can be obtained from the specifications of the light source being tested, or you can directly query the UV light source manufacturer.

I. Choose UV meter according to the type of UV light source

UV Light Source Suggested UV Meter
222nm UVCgermicidal lamp 86125 host+UVC-X2 probe
254nmUVC germicidal lamp 86126C UV light meter
(PC software, split probe)
86127C UV light meter
(PC software, integrated probe)
86125 host+UVC-X0 probe
86125 host+UVCWP-X0 (waterproof)
86129-UVC (online UV light monitor)
UVC LED (260nm, 275nm, etc.) 86125 host+UVCLED-X0 probe
UV plant growth light or reptile lamp 86125 host+UVB-X0 probe
365nmhigh pressure mercury lamp 86125 host+UVA-X0 probe
86129-UVA (online UV light monitor)
UV LED
(365nm, 385nm, 395nm, 405nm, etc.)
86125 host+UVALED-X0 probe
(area light source)
86129-UVALED (online UV light monitor)
86126A UV light meter (ultra-small probe)
86125 host+UVALED-X1 probe
point light source
UV Halogen, gallium, iron, xenon lamps, etc. 86125 host+UVA-X0 / UVA-X1 probe
UV light source in aging box 86125 host+UVA-X2 probe
UV light from the sun 86125 host+UVA-X1/ UVB-X0 probe
Welding arc 86125 host+UVA-X1/ UVB-X0 / UVC-X0 probe
Non-destructive inspection UV lamp 86125 host+UVA-X0 probe
UV LED Nail Lamp 86125 host+UVALED-X3 probe
Blue Light 86125 host+BL-X0 probe

II. Choose UV light meter according to the UV wavelength

Suggested UV Meter UV Spectral Response UV Meter Application
86126C UVC light meter 230-280nm Medical 254nm UVC germicidal
low-pressure mercury lamp
86127C UVC light meter 230-280nm Medical 254nm UVC germicidal
low-pressure mercury lamp
86125 host+UVC-X2 probe 210-250nm 222nm UV germicidal lamps
86125 host+UVC-X0 probe 230-280nm High-power 254nm germicidal mercury lamp
86125 host+UVCWP-X1 probe 230-280nm High-power 254nm UVC germicidal
mercury lamp, 1 meter waterproof
86125 host+UVCLED-X0 probe 240-320nm 260-285nm LED sterilization lamp
86125 host+UVB-X0 probe 280-315nm 297nm, 308nm, 313nm UVB light source
86125 host+UVA-X0probe 315-400nm High pressure mercury lamps
86125 host+UVA-X1probe 260-400nm UVA light source
86125 host+UVA-X2probe 315nm-365nm UVA light aging source
86125 host+UVALED-X0probe 340-420nm UV LED area light source
86125 host+UVALED-X1probe 340-420nm UV LED point light source
86125 host+UVALED-X3 probe 340-420nm Low-power UVA LED light sources
86125 host+BL-X0probe 410nm-490nm LED blue therapy light
86126A UVA light meter 340-420nm UV LED area light source;
small probe or narrow space
86129-UVA online light meter 315-400nm High pressure mercury lamps, online
86129-UVALED online light
meter
340-420nm UV LED area light source,
online
86129-UVC online light meter 230-280nm 254nm UVC germicidal mercury lamp, online


III. Choose UV intensity meter according to the measurement range and test hole diameter

Suggested
UV meter
Power Measuring
Range
Energy/Dosage
Measuring Range
Measuring Aperture
86126C UVC light meter 0-20000 μW/cm²
0-9999999μJ/cm²
10mm
86127C UVC light meter
86125 host+UVC-X0 probe 0-200000 μW/cm²
86125 host+UVC-X2 probe
86125 host+UVCWP-X1 probe
86125+UVCLED-X0 probe
86125 host+UVB-X0 probe
86125 host+UVALED-X3 probe
86125 host+UVA-X1 probe
86125 host+BL-X0 probe 0-2000 mW/cm² 0-9999999mJ/ cm²
86125 host+UVA-X0 probe
86125 host+UVALED-X0 probe 0-20000 mW/cm²
86125 host+UVALED-X1 probe 1mm
86126A UVA light meter 0-9999999mJ/cm² 3mm
86129-UVA online
light meter
0-2000 mW/cm² 0-4×10⁹mJ/ cm² 10mm
86129-UVALED online
light meter
0-20000 mW/cm²
86129-UVC online
light meter
0-200000 μW/cm² 0-4×10⁹μJ/ cm²

IV. FAQ for UV light meters

1. Why there is no traceable calibration certificate for UV LED light meters?

There are no testing standards for LED light source, so we can't provide any traceable reports for the probes or UV intensity meters special for LED light souce (UVCLED-X0 probe, UVALED-X0 probe, UVALED-X1 probe, UVALED-X3 probe, 86126A, 86129-UVA). The data of these probes or UV meters are calibrated based on the manufacturer's standards of Linshang.

We provide factory calibration report for all the UV light intensity meters.

2. Is there a traceable calibration certificate included?

The calibration certificate is not a standard configuration. You need to pay additional calibration costs and wait patiently for the inspection duration of the device.

All the UV light meters, except the probes or UV meters special for LED light source, conform the relevant regulations and standards in China. The UV light intensity meter can pass the inspection of authoritative measurement agency (which has been approved by CNAS and ilac-MRA) and obtain a calibration certificate.

3. What's the value of UV germicidal lamp used in hospitals to be qualified?

The "Chinese Technical Disinfection Technical Specifications" stipulates that the UV light meter is to be placed at a distance of 1 m vertically under a 254 nm germicidal lamp.

The UV light intensity of the new ordinary 30W straight tube must be no less than 100 μW/ cm².

The UV light intensity of the 30W straight tube in use must be no less than 70 μW/ cm².

The UV germicidal lamps with different wattages have various standards, while 30W straight tube is the most commonly used.

The UV light intensity of the germicidal lamp is also related to the ballast. If the ballast of poor quality is used, the qualified UV-C lamp may also be be judged as failing, we recommend that users choose ballasts with good quality, such as Philips.

4. Can these UV meters measure UV wavelength?

Sorry, Linshang UV meters can just measure UV light intensity or UV dose (also UV dosage or energy). Only spectrometers can measure the UV wavelength.

5. Two ways to figure out whether it is a 254nm germicidal lamp (low-pressure mercury lamp)

Option 1:The spectral response range of 86126C or other UV-C probes is 230-280nm, the measured peak wavelength is 254nm, while the UV light meters' response at other wavelengths is very weak. The low-pressure mercury lamp currently known only has a peak wavelength at 254nm. Therefore, if the test data is very low, then the germicidal lamp peak wavelength may not be 254nm or it's unqualified.

Option 2: A spectrometer is needed to measure the spectrum or wavelength of the germicidal lamp.(Unfortuantely this equipment is not provided by Linshang.)

6. UV light meter calibration period and cost

UV meters are generally recommended to be calibrated once a year, however, if they are not used frequently and the UV meters are kept in a dry and clean environment, you can also choose to send them back to the factory for calibration every two years or longer.

The first calibration is free, but the freight cost and VAT/taxes by customs is borne by the client. The cost of subsequent calibration for UV light meter is US$32.