86125 UV Light Meter +UVALED-X3 Probe

Suitable for low-power UVA or UVV LED

Spectral response: 340nm-420nm, calibrated with 395nm LED
Power measuring range: 0-200000μW/cm²
With statistics function, multiple test data are displayed at the same time

User Manual

The 86125 UV light meter is a smart host that can support 9 different types of probes. Its main function is data display. The UVALED-X3 probe is one of the 9 probes equipped with the 86125 host. The probe's spectral response range is 340nm-420nm, calibrated at 395nm. The UV light meter is suitable for detecting the intensity and energy of UVA or UVV light sources, as well as the intensity and energy of low power UVA LED light sources.

Parameters

Parameters UVALED-X3 Probe
Spectral response 340nm-420nm, calibrated with 395nm UV LED
Power measuring range 0 - 200000 μW/cm²
Resolution 0.1 μW/cm²
Energy measuring range 0-9999999 μJ/cm²
Record time 0-99999s
Measuring accuracy ±10%
Sampling speed 6 times/second
Unit μW/cm² (default), mW/cm², W/m²
Test Hole Diameter ¢=10mm
Probe size 39mm ×  32mm × 15mm (L * W * H)
Packing box dimension 24 * 6 * 5 cm (L * W * H)

Features

Dedicated to low-power UV curing light source detection

The intensity of ultraviolet radiation can be measured to the micron level, which is suitable for the intensity and energy detection of low-power UVA light sources

Nail Art Light

Nail Art Light

3D printer

3D printer

Powerful statistics function

The 86125 UV radiometer + UVALED-X3 probe has statistics function, real-time power value, maximum value, minimum value, average value, measurement time and energy value are displayed at the same time

Advanced digital probe technology

Adopting advanced digital probe technology, all data collection and processing are completed inside the probe. Data transmission through digital signals with stronger anti-interference ability can ensure more stable test data

Provide three units to be chosen

Three measurement units can be switched, namely mW/cm², W/cm², W/m². Data conversion between different units does not require manual calculation

FAQ

  1. 1. The difference between UVALED-X0, UVALED-X1 and UVALED-X3 UV probe for 86125 UV light meter host

    1) The test aperture of the UVALED-X0 probe is 10mm, which is mainly used to detect UVA LED plane light sources in curing and exposure industries;

    2) The test aperture of the UVALED-X1 probe is 1mm, which is mainly used to detect UVA LED point light sources in curing and exposure industries;

    3) The UVALED-X3 probe is mainly used for the detection of low-power UVA LED light sources, such as nail art lamps.

  2. 2. What's the working temperature of the UV probes?

    The 9 probes of the 86125 UV light meter are not heat-resistant. The probe can be used for a long time at 80℃ or lower. If the temperature is higher than 80℃, it is recommended to use UV energy meter for UVA light source detection.

  3. 3. Which probe is recommended if I need to measure both mercury lamp and LED light source?

    It is recommended to use the 86125 UV light meter host and UVA-X0, UVALED-X0 probes.

  4. 4. Whether a traceable calibration certificate is included?

    UVA-X0 and UVA-X1 are mainly used to detect 365nm high-pressure mercury lamps, which can be ensured to pass the inspection of the Metrology Institute and obtain a calibration certificate (approved by CNAS and ilac-MRA). The cost of the calibration certificate is about US$180 ; but the UVALED-X0, UVALED-X1 and UVALED-X3 probes are professionally used to detect UVA LED light sources. There is no unified inspection standard for LED light sources, each light source and UV energy meter manufacturer has their own calibration standards. We can provide the factory calibration reports for these three probes special for UV LED, but no calibration certificates can be provided.